Bragg Institute

Specifications

Instrument location: neutron-guide hall, thermal guide TG3

Monochromator

  • Si(400)
  • Resolution: Δd/d < 5 x 10-3 (dependent on combination monochromator-sample)
  • Wavelength: variable between 1 - 2.4 Å due to variable take-off angle and (hkk) reflections from Silicon monochromator (double focusing bend-perfect crystal)
  • angular range: 40°< 2θM < 120°

Sample area

  • gauge volume in the order of 1 mm3
  • flux at sample position 7x106 n cm2s-1 (depends on monochromator reflection and focusing)
  • Maximum sample weight ≤ 1 tonne

Detector

  • 2D, He-filled, 300-300 mm2

Sample environment available

    Ancillary

    Performance

    Quantity

    Cryofurnace-pooled for general use

    6 K - 800 K

    3

    Cryostat


    1.8 K - 300 K

    1

     

    Horizontal load frame

     100 kN

     1