Search
Bragg Institute

Specifications

Instrument location: neutron-guide hall, thermal guide TG3

 

Neutron beamneutronbeam

• Maxwellian distribution with peak intensity at ~1.5 Å
• wavelength range: 1.20-2.85 Å
• take-off angle range: 55-165°
• max beam size: 60x30 mm2


 

 

 

 

 

 

Monochromator
monochromoter

• Double focusing bent perfect crystal
• Si(400) cut with [110] vertically
• Monochromator is made from 9 packets each composed from 13 wafers.
• Wafer dimensions: 270x17x0.91
• Work area: 200x170 mm2
• Thickness: 11.6 mm
•  Available reflections (004), (111), (311), (511), …

 

 

 

 

 

Detectordetector

• DENEX 2D, 3He-filled,
• Total area 300x300 mm2
• Detector coverage: 15°x15° (@ LSD ~ 1000 mm)
• Angular range: 40°< 2θB < 120°
Sample table
• Unique telescopic design
• Maximum sample weight ≤ 1 tonne
• ±250 mm travel for X and Y
• 600 mm travel for Z

 

 

 

 

 


sampletable 

Sample table
  • Unique telescopic design
  • Maximum sample weight ≤ 1 tonne
  • ±250 mm travel for X and Y
  • 600 mm travel for Z

 

 

 

 

 


Instrument performanceinstrumentperformance
  • flux at sample position 7x106 n cm2s-1 (depends on monochromator reflection and focusing)
  • Instrument is optimized for d-spacings ~ 1.1-1.3 Å to ne at ~90°
  • These are the most important reflections in engineering  applications αFe(211), γFe(311), Al(311)
  • For this range resolution is Δd/d ~2 x 10-3
  • Practical examples for time evaluation:
    • Fe(211), gauge volume 0.5x0.5x0.5: t = 10 min to get 50 μstrain accuracy
    • Fe(211), gauge volume 1x1x1: t = 1 min to get 50 μstrain accuracy

 

 

 

 

 

Sample environment available

(1) Horizontal 100kN Instron load frame with different environments
 sampleenvironment

Mirror lamp furnace (up to 1000°C)
mirrorlamp


Environmental chamber (up to 200°C)

environmentalchamber
 
Cold grips (down to -30°C)

coldgrips
 

(2) Big and small Eulerian cradles (goniometers)

Bigeureliansmalleurelian

 

(3) Cybaman manipulator

 

cymbamanmanipulator

(4) Matis laser scanning system

matislaserscanner

Additional general use ancillaries

 

Ancillary

Performance

Quantity

Cryofurnace-pooled for general use 6 K - 800 K 3
Cryostat 1.8 K - 300 K 1
 500C Vacuum Furnace (in house design) 20C to 500C 1
Electric Field Device EF-1 (power supply system) 10kV AC/DC 1
Vacuum Furnace F-1 (ILL type, niobium) 20C - 1600C 1