Instrument location: reactor-beam hall, cold beam CG4 - beam size 50 x 200 mm2 (width x height) at reactor face.
Angular ranges
Monochromator
- Dimensions: 220 mm x 280 mm (width x height)
- PG (002), double focusing
- Cu (111) double focusing
Sample area
- Beam size 65mm x 220mm (W x H)
Distances
- cold source - monochromator 6715mm
- reactor face - monochromator 2300mm
- virtual source - monochromator 2100mm
- monochromator - sample 2100mm (extendable)
Collimators
- a1 (pre-monochromator) 20", 40', 60', open
- a2 (post-monochromator) 20", 40', 60', open
Incident beam filters
- Be cooled (10cm) (Ei < 5 meV)
- PG (8cm) (Ei = 13.7 and 14.7 meV)
- Sapphire (8cm) (Ei > 15meV)
Sample environment available
|
Ancillary |
Performance |
Quantity |
|
Cryofurnace-pooled for general use |
6K - 800K
|
3 |
|
Cryostat |
1.8K-300K
|
1 |
|
Vertical Field Cryomagnet / CCR |
B= 7.4T T=10K-300K |
1 |
|
Eulerian Cradles |
|
2 |