Specifications


Angular ranges

15° < 2θM< 85°
-145° < 2θs< 115°

-110°< 2θA< 110°

Monochromator

200 mm x 200 mm in 9 x 11 segments (width x height)
Continuous horizontal and vertical focussing
Pyrolytic Graphite (002) 24' mosaic with Incident Energy Range: ~ 5 - 60 meV)
Copper (200) 20' mosaic with Incident Energy Range: ~ 14 - 160 meV)

 

Sample area

Beam size at monochromator shielding exit 50mm x 130 mm (W x H)
Flux at sample position is projected to be ~ 2x108 ncm2s-1 at 50 meV

 

Analyser

160 mm x 140 mm in 5 x 7 segments (width x height)
Continuous horizontal and vertical focusing
Pyrolytic Graphite (002) 24' mosaic

 

Soller collimators

Pre-monochromator collimators 15', 30', Open; 90 x 185 mm2 (W x H)
Post-monochromator, pre-analyser and pre-detector, collimators: 20', 40', Open; 50 x 130 mm2 (WxH)

 

Detector

3He detector, (diameter)25mm x 100 mm, p=10 bars
Or (diameter)50mm x 100 mm, p=5 bars

 


Sample environment available

 

AncillaryPerformanceQuantity
Cryofurnace-pooled for general use
 

6K - 800K

3
Cryostat
 
1.8K-300K1
Vacuum Furnace
 
100C-1800C
 
1
Vertical Field Cryomagnet / CCR
 
B= 7.4T T=10K-300K
 
1
Eulerian Cradles
 
 2


 