Ion Beam Induced Current (IBIC)
To function as an effective micro-dosimeter the device must possess an accurately defined charge collection volume.
To measure characteristics of the charge collection of a 10 µm thick device 20 MeV carbon beam was used. Result shows that the magnitude of the charge collection decreases as the ion locations moves further from the n+ region. The distribution is related to the distribution of charge deposited in the Si on Insulator layer see Figures 2a and 2b.
![]() |
| Figure 1: Schematic drawing of micro-dosimeter |
![]() |
| Figure 2: Intensity maps showing a charge collection efficiency- the efficiency decreases as the ion location moves further from n+ region |
![]() |
| Figure 3: Charge collection efficiency of a single device- warmer color indicates higher efficiency |