Skip to main content
A scientist working on the Spatz instrument

Technical Specifications

Wavelength range: Typically use 2 to 19 Å at a disc chopper speed of 25 Hz. 
Wavelength resolution: Δλ/λ from 1% to 12% (5% is used for most experiments)
Qmax: up to 0.3 Å-1 typical for reflection at the solid-liquid and air-solid interface (sample dependent). Detector has a maximum 2θ of 100° for thin-film diffraction options.
Rmin: 10-7 (sample dependent)
Maximum beam size: 50 mm (high) x 30 mm (wide)
Sample geometry: Vertical
Collimation slits:  High precision (< 5μm accuracy), hot pressed and sintered B4C slits
Detector:  DENEX 300TN 3He position sensitive detector 
Off-specular reflection: Yes
Event-mode acquisition: Yes
Polarisation analysis: No (future upgrade) 

 

Instrument paper 

 

A.P. Le Brun, T.-Y. Huang, S. Pullen, A.R.J. Nelson, J. Spedding, S.A. Holt, Spatz: the time-of-flight neutron reflectometer with vertical sample geometry at the OPAL research reactor, Journal of Applied Crystallography (2023) 56(1): 18 – 25 https://doi.org/10.1107/S160057672201086X