Specifications
| Angular ranges |
- 15° < 2θM< 85°
- -145° < 2θs< 115°
- -110°< 2θA< 110°
|
| Monochromator |
- 200 mm x 200 mm in 9 x 11 segments (width x height)
- Continuous horizontal and vertical focussing
- Pyrolytic Graphite (002) 24' mosaic with Incident Energy Range: ~ 5 - 60 meV)
- Copper (200) 20' mosaic with Incident Energy Range: ~ 14 - 160 meV)
|
| Sample area |
- Beam size at monochromator shielding exit 50mm x 130 mm (W x H)
- Flux at sample position is measured to be 1.8x108 ncm2s-1 at 50 meV
|
| Analyser |
- 160 mm x 140 mm in 5 x 7 segments (width x height)
- Continuous horizontal and vertical focusing
- Pyrolytic Graphite (002) 24' mosaic
|
Soller collimators |
- Pre-monochromator collimators 15', 30', Open; 90 x 185 mm2 (W x H)
- Post-monochromator, pre-analyser and pre-detector, collimators: 20', 40', Open; 50 x 130 mm2 (WxH)
|
Detector |
- 3He detector, (diameter)25mm x 100 mm, p=10 bars
- Or (diameter)50mm x 100 mm, p=5 bars
|
| Polarisation analysis |
- 3He polarisers before and after the sample
- Polarisation modes: P||Q and P Perp Q
|
Users Manual
Instrument reference for publications
Rule, KC; Darmann, F; Oste, T; Bartlett, D; Franceschini, F; Berry, A; McGregor, A; Ogrin, A; Ersez, T; Kafes, A; Pangelis, S; Danilkin, S; Stampfl, APJ and Olsen, SR, Recent enhancements and performance gains from upgrades to ANSTO’s thermal neutron instrument TAIPAN and the triple-axis and Be-filter spectrometers, Nucl. Instrum. Methods Phys. Res., Sect. A 901, 140-149(2018) DOI