| Instrument  | Specifications | 
| Monochromator | - Vertically focussing Ge 115
 -   Δd/d > ~ 2 x 10-3 
 
  | 
| Collimators | - Oscillating tertiary collimator reduces critical diameter to 50 mm
 - Must be used with most sample environments
 - Increases minimum 2θ from 10° to 15°
 - Optional secondary collimation
 
  | 
| Sample area | -  Smallest sample weight ~10mg
 -  Typical sample size 10mm3 (restricted due to focusing elements)
 -  1s acquisition for 10mm3 in one shot irreversible experiments
 -  30µs acquisition in stroboscopic mode (reversible experiments)
 
  | 
| Detector | - 120° x 200 mm high 2D detector, 0.125° resolution, maximum angle 160°
 
  | 
Instrument reference
 A. J. Studer, M. E. Hagen and T. J. Noakes, Physica B 385-386, 1013-1015 (2006).